Isotropic antennas
|
R&S®TSEMF-B3
|
R&S®TSEMF-B1
|
R&S®TSEMF-B2
|
Measurement principle
|
isotropic reception due to orthogonally arranged
antenna elements that are electronically switched
|
Frequency range
|
9 kHz to 200 MHz
|
30 MHz to 3 GHz
|
700 MHz to 6 GHz
|
Minimum detectable field strength
|
approx. 1 mV/m
|
Maximum field strength
|
≥100 V/m
|
≥230 V/m
|
≥230 V/m
|
Isotropy
|
≤ ±1.37 dB
|
≤ ±2.1 dB,
±1.0 dB (f = 900 MHz),
±1.7 dB (f = 1800 MHz)
|
≤ ±2.5 dB (0.6 GHz to 2 GHz),
≤ ±2.2 dB (2 GHz to 3.6 GHz),
≤ ±3.0 dB (3.6 GHz to 6 GHz)
|
Mechanical design
|
radome protection against mechanical damage and
environmental hazards
|
Antenna factor
|
individual calibration data, saved on USB dongle and/or
CD
|
Axis switching
|
RF solid state switch
|
Connecting cables
|
direct connection to spectrum analyzer (extension cable
(length: 8 m) see accessories)
|
integrated cable (length: 2 m) ferritebeaded
(extension cable (length: 8 m) see accessories)
|
integrated cable (length: 2 m, extension cable
(length: 8 m) see accessories))
|
RF connector
|
N male
|
Connector, control line
|
7-pin connector (binder) for direct connection to
R&S®FSH power
sensor port, adapter cable for connection to notebook or spectrum analyzer
with USB output
|
Tripod adapter
|
?-" thread, quick connector for antenna
|
Requirements for notebook (not supplied)
|
MS Windows7 or Windows8 operating system, free hard
disk space: min. 5 Gbyte
display resolution: min. 800 × 600 pixel, two USB interfaces
interface to test instrument (depending on analyzer: GPIB, LAN, USB or
FireWire)
recommended application: Excel 2007, 2010, 2013
|
Equipment supplied
|
isotropic sensor with connecting cable operator’s guide
for R&S®RFEX and
R&S®RFEX-Fast
measurement software (on CD), tripod adapter, USB adapter cable, transit case
(R&S®TS-EMF with at
least one isotropic antenna)
|
Expanded measurement uncertainty R&S®TS-EMF with
R&S®FSH (95 %
confidence level (k = 2))
|
≤ ±2.5 dB
|
≤ ±3.3 dB,
±2.3 dB at 0.9 GHz,
±2.9 dB at 1.8 GHz
|
≤ ±3.1 dB (0.7 GHz to 1 GHz),
≤ ±3.3 dB at (1 GHz to 3.6 GHz),
≤ ±3.7 dB at (3.6 GHz to 6 GHz)
|
Expanded measurement uncertainty R&S®TS-EMF with
R&S®FSL (95 %
confidence level (k = 2))
|
≤ ±2.4 dB
|
≤ ±3.2 dB,
±2.2 dB at 0.9 GHz,
±2.9 dB at 1.8 GHz
|
≤ ±2.9 dB (0.7 GHz to 1 GHz),
≤ ±3.2 dB at (1 GHz to 3.6 GHz),
≤ ±3.6 dB at (3.6 GHz to 6 GHz)
|
Power supply
|
via spectrum analyzer or laptop
|
Ambient conditions
|
–10 °C to +50 °C, safety class IP54 (see user
manual for details)
|
Weight (incl. cable)
|
0.85 kg
1.87 lb
|
1.3 kg
2.87 lb
|
0.95 kg
2.09 lb
|
Dimensions (L × ?)
|
550 mm × 146 mm
21.7 in. × 5.75 in.
|
475 mm × 170 mm
18.7 in. × 6.69 in.
|
415 mm × 87 mm
16.3 in. × 3.43 in.
|